LPS60-M Medical Series WebSheet
- Total Max Power: 80 W
- Total Max Current: 16 A
- Input Voltage: 85 V to 264 V
- Output Voltage: 12 V to 24 V
- Outputs: Single
- Size (L x W x H): 5.000" x 3.000" x 1.650"
- Warranty: 2 years
- Operating Temperature: -20 to +70
Download Datasheet
Special features
- Medical safety approvals
-
3" x 5" footprint
-
Built-in EMI filter
-
Overvoltage protection
-
Adjustable output
-
Low output ripple
-
Remote sense
-
Universal input
-
110 KHz switching frequency
-
LPX60 enclosure kit available
-
Overload protection
Electrical specs
Input
- Frequency: 47-440 Hz
- EMI filter: FCC Class B conducted CISPR 22 Class B conducted E
- Inrush current: <18 A peak @ 115 VAC; <36 A peak @ 230 VAC,
- Input range: 85-264 VAC 120-300 VDC
- Input current: 1.5 A max. (RMS) @ 115 VAC
- Efficiency: 70% typical at full load
- Safety ground leakage current: <0.5 mA @ 50/60 Hz, 264 VAC input
Output
- Remote sense: Compensates for 0.5 V lead drop min.
- Hold-up time: 20 ms @ 60 W load, 115 VAC nominal line
- Adjustment range: -5, +10% minimum
- Overload protection: Short circuit protection on all outputs.
- Maximum power: 60 W for convection; 80 W with 30 CFM forced air
- Overvoltage protection: 5 V output: 5.7-6.7 VDC. Other outputs 10% to 25%
LPS60-M Medical models