LPT60-M Medical Series WebSheet
- Total Max Power: 80 W
- Total Max Current: 8 A
- Input Voltage: 852 V to 264 V
- Output Voltage: -15 V to 19.5 V
- Outputs: Triple
- Size (L x W x H): 5.000" x 3.000" x 1.650"
- Warranty: 2 years
- Operating Temperature: -20 to +70
Download Datasheet
Medical safety approvals
Universal input
3" x 5" footprint
Remote sense on main output
Built-in EMI filter
Adjustable main output
LPX60 enclosure kit available
Low output ripple
Overload protection
110 KHz switching frequency
Overvoltage protection
Electrical specs
Input
- Inrush current: <18 A peak @ 115 VAC; <36 A peak @ 230 VAC
- Frequency: 47-440 Hz
- Input current: 1.5 A max. (RMS) @ 115 VAC
- Safety ground leakage current: <0.5 mA @ 50/60 Hz, 264 VAC input
- EMI filter: FCC Class B conducted CISPR 22 Class B conducted E
- Efficiency: 70% typical at full load
- Input range: 85-264 VAC 120-300 VDC
Output
- Hold-up time: 20 ms @ 60 W load, 115 VAC nominal line
- Overload protection above peak rating: Short circuit protection on all outputs
- Cross regulation: ±2% on output 1; ±5% on outputs 2,3
- Remote sense: Compensates for 0.5 V lead drop min.
- Adjustment range: -5, +10% minimum
- Maximum power: 60 W for convection; 80 W with 30 CFM
LPT60-M Medical models
| Output(s) |
Model No |
Stock Check |
Evaluation |
| 5.5 V @ 8 A, 14.88 V @ 3.5 A, -12 V @ 1 A |
LPT62-M |
Check |
Request |
| 5.5 V @ 8 A, 19.5 V @ 3.3 A, -15 V @ 1 A |
LPT63-M |
Check |
Request |