Network Power Divisions:
Live Chat     Call+1 888 412 7832     EmailEmail Us

LPT60-M Medical Series WebSheet

LPT60-M Medical
  • Total Max Power: 80 W
  • Total Max Current: 8 A
  • Input Voltage: 852 V to 264 V
  • Output Voltage: -15 V to 19.5 V
  • Outputs: Triple
  • Size (L x W x H): 5.000" x 3.000" x 1.650"
  • Warranty: 2 years
  • Operating Temperature: -20 to +70
  • Download Datasheet

Medical safety approvals
Universal input
3" x 5" footprint
Remote sense on main output
Built-in EMI filter
Adjustable main output
LPX60 enclosure kit available
Low output ripple
Overload protection
110 KHz switching frequency
Overvoltage protection

Electrical specs

Input

  • Inrush current: <18 A peak @ 115 VAC; <36 A peak @ 230 VAC
  • Frequency: 47-440 Hz
  • Input current: 1.5 A max. (RMS) @ 115 VAC
  • Safety ground leakage current: <0.5 mA @ 50/60 Hz, 264 VAC input
  • EMI filter: FCC Class B conducted CISPR 22 Class B conducted E
  • Efficiency: 70% typical at full load
  • Input range: 85-264 VAC 120-300 VDC

Output

  • Hold-up time: 20 ms @ 60 W load, 115 VAC nominal line
  • Overload protection above peak rating: Short circuit protection on all outputs
  • Cross regulation: ±2% on output 1; ±5% on outputs 2,3
  • Remote sense: Compensates for 0.5 V lead drop min.
  • Adjustment range: -5, +10% minimum
  • Maximum power: 60 W for convection; 80 W with 30 CFM

LPT60-M Medical models

Output(s) Model No Stock Check Evaluation
5.5 V @ 8 A, 14.88 V @ 3.5 A, -12 V @ 1 A LPT62-M Check Request
5.5 V @ 8 A, 19.5 V @ 3.3 A, -15 V @ 1 A LPT63-M Check Request